Ambient oxygen concentrations resulting from use of the Lund University Cardiopulmonary Assist System (LUCAS) device during simulated cardiopulmonary resuscitation.
Resuscitation · 2007-05-15 · Method & validation
Abstract
INTRODUCTION: Supplementary oxygen is used routinely during cardiopulmonary resuscitation (CPR). High ambient oxygen levels from ventilation circuits have caused fatal fires and explosions. The Lund University Cardiopulmonary Assist System (LUCAS) device is driven by > 70 l min(-1) oxygen which is also likely to increase ambient oxygen concentrations and cause similar risk of fire and explosion. This study used simulated CPR with a LUCAS device to measure resulting ambient oxygen concentrations and assess safety of the device. MATERIAL AND METHODS: Simulated manikin CPR using a LUCAS device was performed outdoors, inside an ambulance (ventilation off and ventilation on full power), and in a resuscitation bay. Ambient oxygen concentrations were measured over the apical and sternal defibrillation sites and midway between the two, at the head and 1m horizontally above the head. Recordings were made for 5 min when the LUCAS device was turned on and for a further 5 min when turned off. RESULTS: Ambient oxygen concentration increased quickly in all four scenarios. Peak oxygen levels over the chest were highest in the resuscitation bay (36.7%) and lowest in the ambulance with ventilation on full power (33.8%). Oxygen levels decreased to baseline within 5 min of turning off the LUCAS device. CONCLUSION: The use of oxygen to drive the LUCAS device results in a rapid increase in ambient oxygen concentration to levels likely to risk injury or death from fire. Ambulance services and hospitals using the device must be alerted to these dangers immediately.
Abstract © the authors or publisher, via PubMed. Reuse and removal
Tags
Other drugs · Mechanical CPR · Oxygen
Related articles
- Ambient oxygen enrichment during cardiopulmonary resuscitation: an overestimated risk? - Insights from a manikin study. (Scand J Trauma Resusc Emerg Med, 2026)
- Confirming the Clinical Safety and Feasibility of a Bundled Methodology to Improve Cardiopulmonary Resuscitation Involving a Head-Up/Torso-Up Chest Compression Technique. (Crit Care Med, 2019)
- Use of the Lund University Cardiopulmonary Assist System in the MD902: a fire safety assessment. (Emerg Med J, 2007)
- Continuous chest compressions with a simultaneous triggered ventilator in the Munich Emergency Medical Services: a case series. (Ger Med Sci, 2019)
- The use of supplemental oxygen during cardiopulmonary resuscitation - a systematic review of the literature. (Resusc Plus, 2026)
- Novel Pharyngeal Oxygen Delivery Device Provides Superior Oxygenation during Simulated Cardiopulmonary Resuscitation. (Prehosp Disaster Med, 2024)