Usability and performance of a wearable tele-echography robot for focused assessment of trauma using sonography.
Medical engineering & physics · 2012-05-20 · Method & validation
Abstract
Focused Assessment with Sonography for Trauma (FAST) is widely used as a first lifesaving step for patients suffering from internal bleeding. Because it may take a long time to transport such patients to a hospital, a wearable and portable tele-echography robot that a paramedic can attach to the patient has been developed. In the current study, experiments were conducted to evaluate the usability and performance of attached FAST. The proposed robot must be attached to 4 areas to perform FAST. The time required for attachment and the positions of attachment completed by 9 non-medical staff members, as well as the time it took for the FAST to reach a medical doctor, were measured. The echo images obtained when the patient's body was in motion were evaluated by a medical doctor. The robot could be attached to all 4 areas within approximately 5min, and the maximum gap was 4.8cm. This indicates that a paramedic who has received training in emergency medical care should be able to attach the robot to a patient quickly and accurately. Additionally, it was confirmed that the robot could be used to complete FAST under a doctor's control within 9min and that the extracted echo images were suitable for FAST. A comparison of the results with current ambulance transportation time confirmed that FAST could be completed approximately 14min before the patient reached the hospital. The results of the current study indicate that the robot is worth using, is suitable for FAST, and will be effective in emergency medical care.
Abstract © the authors or publisher, via PubMed. Reuse and removal
Tags
Point-of-care ultrasound · Telemedicine
Related articles
- Emergency Life-Saving Technician-Performed Focused Assessment with Sonography for Trauma in Moving Ambulances: A Pilot Innovation Report. (Prehosp Disaster Med, 2026)
- Impact of Telemedicine on Extended Focused Assessment With Sonography for Trauma Performance and Workload by Critical Care Transport Personnel. (Air Med J, 2023)
- Clinical applicability of real-time, prehospital image transmission for FAST (Focused Assessment with Sonography for Trauma). (J Telemed Telecare, 2013)
- Evaluation of a wearable tele-echography robot system: FASTele in a vehicle using a mobile network. (Annu Int Conf IEEE Eng Med Biol Soc, 2011)
- Tele-ultrasound and paramedics: real-time remote physician guidance of the Focused Assessment With Sonography for Trauma examination. (Am J Emerg Med, 2011)
- Satellite and mobile wireless transmission of focused assessment with sonography in trauma. (Acad Emerg Med, 2003)