T-piece or self inflating bag for positive pressure ventilation during delivery room resuscitation: an RCT.
Resuscitation · 2015-01-28 · Randomised trial
Abstract
OBJECTIVE: To compare the duration of positive pressure ventilation (PPV) during delivery room resuscitation in neonates resuscitated with self-inflating bag (SIB) and T-piece resuscitator (TPR). DESIGN: Randomized control trial. SETTING: Delivery room and neonatal intensive care unit of a tertiary care center in northern India. PATIENTS: Consecutively born neonates more than 26 weeks of gestation requiring PPV at birth. INTERVENTION: Eligible neonates were randomized to two groups, SIB and TPR. OUTCOME MEASURES: Duration of PPV, intubation rates in delivery room, incidence of respiratory distress, need for mechanical ventilation during first 48h and its duration, need for surfactant replacement therapy and mortality during NICU stay. RESULTS: Fifty neonates received PPV with a SIB and 40 received PPV with a TPR. The mean (SD) birth weight and gestational age of neonates in SIB and TPR groups were 2264 (872) and 2065 (814)g; 35.1 (3.6) and 34.3 (3.7) weeks, respectively. The median (IQR) duration of PPV in delivery room was significantly less in TPR group as compared to SIB; 30 (30-60)s vs. 60 (30-90)s, respectively; (p<0.001). A higher proportion of neonates required delivery room intubation in SIB group as compared to TPR group (34% vs. 15%, p=0.04). In the TPR group, a higher proportion of neonates could be resuscitated with room air only (72.5% vs. 38%, p=0.001). Other outcomes were comparable in the two groups. Similar findings were observed in neonates <34 weeks, except that fewer neonates resuscitated with TPR required invasive ventilation (31.6% vs. 77.8%, p=0.008). CONCLUSION: Use of TPR during delivery room resuscitation resulted in shorter duration of PPV and lesser rates of intubation as compared to SIB. More infants in this group could be resuscitated with room air only (CTRI/2010/091/002946).
Abstract © the authors or publisher, via PubMed. Reuse and removal
Tags
Bag-valve-mask ventilation · Mortality & survival · Neonates
Related articles
- Ventilation devices for neonatal resuscitation at birth: A systematic review and meta-analysis. (Resuscitation, 2023)
- Comparison of devices for newborn ventilation in the delivery room. (J Pediatr, 2014)
- Effective ventilation and temperature control are vital to outborn resuscitation. (Prehosp Emerg Care, 2004)
- A Randomized Trial of Laryngeal Mask Airway in Neonatal Resuscitation. (N Engl J Med, 2020)
- A systematic review of mortality and severe morbidities following newborn chest compressions at birth. (Resuscitation, 2026)
- Chest Compression Superimposed with Sustained Inflation or 3:1 Compression/Ventilation Ratio During Neonatal Cardiopulmonary Resuscitation in the Delivery Room: A Systematic Review and Meta-Analysis. (Children (Basel), 2025)