Differences in Automated External Defibrillator Types in Out-of-Hospital Cardiac Arrest Treated by Police First Responders.
Journal of Cardiovascular Development and Disease · 2023-04-27 · Observational study
Abstract
Background: Police first responder systems also including automated external defibrillation (AED) has in the past shown considerable impact on favourable outcomes after out-of-hospital cardiac arrest (OHCA). While short hands-off times in chest compressions are known to be beneficial, various AED models use different algorithms, inducing longer or shorter durations of crucial timeframes along basic life support (BLS). Yet, data on details of these differences, and also of their potential impact on clinical outcomes are scarce. Methods: For this retrospective observational study, patients with OHCA of presumed cardiac origin and initially shockable rhythm treated by police first responders in Vienna, Austria, between 01/2013 and 12/2021 were included. Data from the Viennese Cardiac Arrest Registry and AED files were extracted, and exact timeframes were analyzed. Results: There were no significant differences in the 350 eligible cases in demographics, return of spontaneous circulation, 30-day survival, or favourable neurological outcome between the used AED types. However, the Philips HS1 and -FrX AEDs showed immediate rhythm analysis after electrode placement (0 [0-1] s) and almost no shock loading time (0 [0-1] s), as opposed to the LP CR Plus (3 [0-4] and 6 [6-6] s, respectively) and LP 1000 (3 [2-10] and 6 [5-7] s, respectively). On the other hand, the HS1 and -FrX had longer analysis times of 12 [12-16] and 12 [11-18] s than the LP CR Plus (5 [5-6] s) and LP 1000 (6 [5-8] s). The duration from when the AED was turned on until the first defibrillation were 45 [28-61] s (Philips FrX), 59 [28-81] s (LP 1000), 59 [50-97] s (HS1), and 69 [55-85] s (LP CR Plus). Conclusion: In a retrospective analysis of OHCA-cases treated by police first responders, we could not find significant differences in clinical patient outcomes concerning the respective used AED model. However, various differences in time durations (e.g., electrode placement to rhythm analysis, analysis duration, or AED turned on until first defibrillation) along the BLS algorithm were seen. This opens up the question of AED-adaptations and tailored training methods for professional first responders.
Abstract © the authors or publisher, via PubMed. Reuse and removal
Tags
Automated external defibrillator · Fire service & police · Mortality & survival · Quality of life & function
Related articles
- Out-of-hospital cardiac arrest outcomes when law enforcement arrives before emergency medical services. (Resuscitation, 2024)
- Sudden Cardiac Arrests in the Polish Tatra Mountains: A Retrospective Study. (Wilderness Environ Med, 2023)
- The association of fire or police first responder initiated interventions with out of hospital cardiac arrest survival. (Resuscitation, 2022)
- The impact of a high-quality basic life support police-based first responder system on outcome after out-of-hospital cardiac arrest. (PLoS One, 2020)
- Association of Public Health Initiatives With Outcomes for Out-of-Hospital Cardiac Arrest at Home and in Public Locations. (JAMA Cardiol, 2017)
- Association of Bystander and First-Responder Intervention With Survival After Out-of-Hospital Cardiac Arrest in North Carolina, 2010-2013. (JAMA, 2015)