Intrathoracic pressure regulation during cardiopulmonary resuscitation: a feasibility case-series.
Resuscitation · 2012-08-17 · Observational study
Abstract
AIM OF THE STUDY: Intrathoracic pressure regulation (IPR) is a novel, noninvasive therapy intended to increase cardiac output and blood pressure in hypotensive states by generating a negative end expiratory pressure of -12 cm H2O between positive pressure ventilations. In this first feasibility case-series, we tested the hypothesis that IPR improves End tidal (ET) CO2 during cardiopulmonary resuscitation (CPR). ETCO2 was used as a surrogate measure for circulation. METHODS: All patients were treated initially with manual CPR and an impedance threshold device (ITD). When IPR-trained medics arrived on scene the ITD was removed and an IPR device (CirQLATOR™) was attached to the patient's advanced airway (intervention group). The IPR device lowered airway pressures to -9 mmHg after each positive pressure ventilation for the duration of the expiratory phase. ETCO2, was measured using a capnometer incorporated into the defibrillator system (LifePak™). Values are expressed as mean ± SEM. Results were compared using paired and unpaired Student's t test. p values of <0.05 were considered statistically significant. RESULTS: ETCO2 values in 11 patients in the case series were compared pre and during IPR therapy and also compared to 74 patients in the control group not treated with the new IPR device. ETCO2 values increased from an average of 21 ± 1 mmHg immediately before IPR application to an average value of 32 ± 5 mmHg and to a maximum value of 45 ± 5mmHg during IPR treatment (p<0.001). In the control group ETCO2 values did not change significantly. Return of spontaneous circulation (ROSC) rates were 46% (34/74) with standard CPR and ITD versus 73% (8/11) with standard CPR and the IPR device (p<0.001). CONCLUSIONS: ETCO2 levels and ROSC rates were significantly higher in the study intervention group. These findings demonstrate that during CPR circulation may be significantly augmented by generation of a negative end expiratory pressure between each breath.
Abstract © the authors or publisher, via PubMed. Reuse and removal
Tags
CPR adjuncts · Capnography · Mortality & survival
Related articles
- Focused echocardiography, end-tidal carbon dioxide, arterial blood pressure or near-infrared spectroscopy monitoring during paediatric cardiopulmonary resuscitation: A scoping review. (Resusc Plus, 2021)
- The effect of alternative methods of cardiopulmonary resuscitation - Cough CPR, percussion pacing or precordial thump - on outcomes following cardiac arrest. A systematic review. (Resuscitation, 2021)
- Impact of impedance threshold devices on cardiopulmonary resuscitation: a systematic review and meta-analysis of randomized controlled studies. (Crit Care Med, 2008)
- Effect of Interposed Abdominal Compression on Cardiopulmonary Resuscitation Outcomes; a Randomized Clinical Trial. (Arch Acad Emerg Med, 2022)
- Confirming the Clinical Safety and Feasibility of a Bundled Methodology to Improve Cardiopulmonary Resuscitation Involving a Head-Up/Torso-Up Chest Compression Technique. (Crit Care Med, 2019)
- End-tidal carbon dioxide output in manual cardiopulmonary resuscitation versus active compression-decompression device during prehospital quality controlled resuscitation: a case series study. (Emerg Med J, 2018)